SA-260MB

Autoclave

The Series represents Sturdy’s all purpose, top-of-line tabletop autoclave with versatility and flexibility to meet almost any situation that arises. This model uses an LCD panel along with 3 navigation buttons (up/down/return), which allow you to configure almost everything using the screen - this minimizes the buttons on the panel and allows the series more flexibility and functionality.

  • Double water tanks
    LCD Panel, Friendly use
    Pre & post vacuum
    Prion preset program
    Leakage, Helix, B.D. diagnostics
    Vacuum based tubal sterilization
    Tank can be filled manually or piped-in (filtered water)
    Process evaluation system
    Customization program
    Chamber size: 24 L

Double water tanks
LCD Panel, Friendly use
Pre & post vacuum
Prion preset program
Leakage, Helix, B.D. diagnostics
Vacuum based tubal sterilization
Tank can be filled manually or piped-in (filtered water)
Process evaluation system
Customization program
Chamber size: 24 L

Specification

Model SA-260MB
Chamber Capacity (L) 24
Maximum Instrument Length (mm) 350
Maximum Load (unwrapped, solid) (g) 5,000
Maximum Load (wrapped) (g) 1,500
External Dimensions (mm) 553 (W) ×440 (H) × 665 (D) 
Chamber Size (mm) Ø260 mm × 450 mm (D)
Gross Weight (kg) 54
Voltage/Wattage (Heater) 230V AC, 50/60Hz, 12A
Heater 1800W for main heater. 870W for band heater
Fuses 15A × 2, No Fuse (circuit) Breaker
Water Reservoir Capacity (ml) 4200
Water Capacity per Cycle (ml) 1270
Sterilization Temperature (°C) 105 – 135
Working Environment Indoor use;
Under 1,000m (altitude);
Temperature 5°C to 40°C;
Relative Humidity 80%RH@31°C to Relative Humidity 50%RH@40°C;
Voltage fluctuation ±10 %;
Transient overvoltages category II;
Pollution degree 2
Transportation Conditions -10°C to 70°C, 10%RH to 90%RH
Storage Conditions -10°C to 50°C, 10%RH – 70%RH
Designed Temperature(°C) 142
Designed Pressure 2.76 kgf/cm² (2.7 bar)
Over Pressure Protection 2.55 kgf/cm² (2.5 bar)
Air Filter Efficiency  μm≦0.3
Over Pressure Indication Yes
Over Temperature Indication Yes
Water Level Indication Yes
Door Lock Indication Micro switch sensor with warning LCD
Pressure Display Analog pressure gauge, LCD display
Function Display LCD
Sterilization Program Unwrapped 121°C
Wrapped 121°C
Unwrapped 134°C
Wrapped 134°C
PRION
LIQUID 105-135°C(Optional)
Customization 105-135°C
Test Program Leakage test,
Helix test,
Bowie-Dick test
Dry Program 1-60 minutes
Others Function Cancel,
Emergency.
Sterilization process recording,
Auto add water,
Real-time Printer,
Cycle counter,
Next Service cycles remind,
Unit Setting for Pressure and Temperature,
Date and time setting
Calibration Mode/Engineering Mode
Printer Thermal Printer
Max. capacity of SD card  SD/HC (Max. 32GB)

Download

GLOBAL

Cookies

We use cookies to optimize this website and continuously update it according to your needs.Please click on "Cookies" for detailed information on the use of cookies on this website. If you click "Yes, I agree", cookies will be activated. If you do not want cookies to be activated, you can opt out here. The settings can be changed at any time under "Cookies".

Yes, I agree.